iCup Dx14 Drug Screen Cup(AMT Item Number/ AMT Description): X11-I-DXA-2147-011/ "iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC150)+(MOP300/MET500)+(AMP500/BZO300)+ (BAR300/MTD300)+(BUPG10/TCA1000)+(MDMA500/OXY100)+(PCP25/PPX300) 5 Adulteration Tests: Cr, Ni, pH, Bl, S.G.", X11-I-DX-1147-022 /"iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC150)+(MOP300/MET500)+(AMP500/BZO300)+ (BAR300/MTD300)+(BUPG10/TCA1000)+(MDMA500/OXY100)+(PCP25/PPX300)"
Product Description
iCup Dx14 Drug Screen Cup(AMT Item Number/ AMT Description): X11-I-DXA-2147-011/ "iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC150)+(MOP300/MET500)+(AMP500/BZO300)+ (BAR300/MTD300)+(BUPG10/TCA1000)+(MDMA500/OXY100)+(PCP25/PPX300) 5 Adulteration Tests: Cr, Ni, pH, Bl, S.G.", X11-I-DX-1147-022 /"iCup" Dx 14 Drug Screen Cup 14 Drug Detection: (THC50/COC150)+(MOP300/MET500)+(AMP500/BZO300)+ (BAR300/MTD300)+(BUPG10/TCA1000)+(MDMA500/OXY100)+(PCP25/PPX300)"
Reason for Recall
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Details
- Recalling Firm
- Ameditech Inc
- Units Affected
- 136,925 units
- Distribution
- US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
- Location
- San Diego, CA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11