Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.
Product Description
Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.
Reason for Recall
Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.
Details
- Recalling Firm
- Pentax of America Inc
- Units Affected
- US: 118 units
- Distribution
- Nationwide
- Location
- Montvale, NJ
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11