Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each pad is employed for testing one assay item by visually or instrumentally reading the color change of the pad and comparing with the corresponding blocks on a color chart. The Healgen Series Reagent Strips for Urinalysis provides tests for Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Ascorbic Acid and Leukocytes in Urine.

Reported: October 9, 2024 Initiated: August 22, 2024 #Z-3303-2024

Product Description

Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each pad is employed for testing one assay item by visually or instrumentally reading the color change of the pad and comparing with the corresponding blocks on a color chart. The Healgen Series Reagent Strips for Urinalysis provides tests for Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Ascorbic Acid and Leukocytes in Urine.

Reason for Recall

The reason for this recall is Healgen Scientific Ltd. has determined that their Urinary Reagent Strip products have shown reduced sensitivity, or potential false negative results for the detection of urinary protein by way of uneven color fading on the total protein test pad. This could lead to a delay in the identification of potential kidney diseases, their evaluation, and/or other conditions associated with elevated protein levels in urine and subsequently a delay in patient treatment.

Details

Recalling Firm
Healgen Scientific Llc
Units Affected
1,636,994 Tests
Distribution
U.S. Nationwide distribution in the states of CA, GA and TX.
Location
Houston, TX

Frequently Asked Questions

What product was recalled?
Healgen Series Reagent strips for Urinalysis are composed of several color pads aligned on a test strip. Each pad is employed for testing one assay item by visually or instrumentally reading the color change of the pad and comparing with the corresponding blocks on a color chart. The Healgen Series Reagent Strips for Urinalysis provides tests for Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Ascorbic Acid and Leukocytes in Urine.. Recalled by Healgen Scientific Llc. Units affected: 1,636,994 Tests.
Why was this product recalled?
The reason for this recall is Healgen Scientific Ltd. has determined that their Urinary Reagent Strip products have shown reduced sensitivity, or potential false negative results for the detection of urinary protein by way of uneven color fading on the total protein test pad. This could lead to a delay in the identification of potential kidney diseases, their evaluation, and/or other conditions associated with elevated protein levels in urine and subsequently a delay in patient treatment.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 9, 2024. Severity: Moderate. Recall number: Z-3303-2024.